How to Use a Lancing Device: 6 Steps and Depth Settings
Knowing how to use a lancing device correctly is mostly a matter of six steps — load, prime, set the depth, fire, eject, clean — but the order matters less than one rule that sits above all of them: the pen is a single-patient instrument. We manufacture both the disposable safety lancets that replace these pens in clinical settings and the lancets that feed them, so this guide covers the mechanics and the rules together. Six numbered steps, a depth-setting table, the cleaning schedule regulators expect to see documented, and the five mistakes that keep showing up in outbreak investigations.
Short answer: insert a fresh lancet until it clicks, twist off its protective cap, prime the device by pressing or pulling the plunger until it locks, turn the depth dial — start low and move up until the drop is adequate — press the cap flat against the side of the fingertip and fire, then eject the used lancet into a sharps container and disinfect the pen weekly with 70% isopropanol. Never share the device, even with a new lancet.
Contents
- Who this guide is for, and the one rule above the steps
- Step 1: how to load a lancet into a lancing device
- Step 2: priming — cocking the mechanism
- Step 3: choosing a depth setting on the dial
- Steps 4 and 5: firing, and what to do with the drop
- Step 6: ejecting the lancet safely
- Cleaning and disinfecting: the schedule and the method
- How to use a lancing device safely: five recurring mistakes
- Buying note: what to check before a facility adopts a pen
- Frequently asked questions
Who this guide is for, and the one rule above the steps
An adjustable lancing device is a reusable holder that fires disposable lancets. The lancet is sterile and single-use. The pen is not sterile, is not disposable, and is not cleanable to a degree that lets two people share it.
That last sentence is not our opinion; it is the regulatory position. In the United States, a reusable lancing device cleared for a single patient is a Class II device. One cleared or used for multiple patients has no lawful market path at all: FDA’s 2021 final order placed multi-patient reusable devices in Class III, requiring premarket approval, and the submission deadline for those devices expired in May 2024. A pen used between patients today is, functionally, an unapproved device.
The single-patient rule, in plain terms. One pen, one person, for life. It is not shared between family members. It is not used by a nurse to test several patients, even with a fresh lancet in it each time. For assisted blood draws — where a carer operates the device — FDA guidance says the carer should use a single-use safety lancet that permanently retracts, not the patient’s reusable pen. The reason is documented: in long-term-care hepatitis B outbreaks, investigators found transmission continued even though the blade was changed for every patient, because the contamination lived on the reusable base.
Everything below assumes that rule. If your setting is a hospital ward or a screening programme, this device is the wrong category entirely — that is a single-use safety lancet job, and our safety lancet usage guide covers it instead.
Step 1: how to load a lancet into a lancing device
Remove the cap from the pen. Take a sterile lancet from its packaging — check the expiry and that the packaging is intact; a torn foil pouch means a non-sterile needle, and there is no way to sterilise it at the point of care.
Insert the lancet into the holder at the base of the device, pushing until it clicks into place. The click is the seating mechanism, not a formality: an unseated lancet can fire at the wrong depth or fail to eject. Then twist off the lancet’s protective cap and set it aside — you will want it again at ejection time on many devices.
Two details cause most loading problems:
Twist-off caps: remove the whole cap, not half of it
Twist lancets have a rigid cap over the needle. If it is only partially removed, the needle cannot reach full extension and the puncture will be too shallow — a common cause of “the device is not working” complaints that are actually cap problems. Pull the cap off in one motion and inspect: the needle tip should be fully exposed and straight.
Which lancets fit
Pens take two feeding formats. Twist-top plastic lancets — usually 28G to 30G — are the commonest, and the stainless strip lancets used in many clinical and laboratory pens are the other. The formats are not interchangeable, and the pen’s instructions for use name what it takes. Compatibility details between the two formats, and why “flat” on a box can mean either, are covered in our twist-off versus flat lancet comparison. Match the gauge to the test, not the other way round: 30G at 1.5 mm covers routine glucose; deeper or wider needs sit in the 21G to 23G range, as set out in our gauge and depth chart.
Step 2: priming — cocking the mechanism
Priming loads the spring. On most pens this is done by pressing the plunger button until it locks, or by pulling the adaptive cap or a slide back until it catches. The ready state is usually visible — a colour change on a release button, or a visible gap that closes on firing.
Two rules here prevent the commonest failures:
- Prime only when you are ready to puncture. A cocked device fires if the release button is pressed against anything — a pocket, a bedsheet, another patient’s tray. Prime at the point of use, fire within seconds, and never carry a primed pen.
- Do not double-prime. If the device is primed and you need to reposition, leave it cocked; cycling the plunger without firing wastes the stroke and, on some designs, can partially advance the needle.
Step 3: choosing a depth setting on the dial
The dial on the cap selects how far the needle travels. Settings are discrete steps — five to eleven positions on mainstream pens, each a fixed increment — not a continuous adjustment. The numbers are indexed by skin condition and blood requirement, not by patient age.
| Situation | Starting setting | Reasoning |
|---|---|---|
| First use, any patient | Lowest setting | Adjust up only if the drop is inadequate. You cannot un-stick a bruise. |
| Thin or tender skin | Low | Thin skin reaches the capillary loops at shallow depths. |
| Thick or calloused skin | Mid to high | The callus absorbs depth before the needle reaches the papillary dermis. |
| Standard glucose test (0.6–1 µL) | Low to mid | One drop is 30–50 µL. Most meters need under 1. |
| Larger volume tests | Mid to high | Move up a step at a time until flow is adequate, then stop. |
| Cold hands, poor perfusion | Warm first, then mid | Perfusion fixes more low-flow problems than depth does. Warm the site up to 40–42 °C for 3–5 minutes. |
Start low, adjust up, and record where you ended
The routine that works for self-testing: begin at the lowest setting, fire once, and if the drop is inadequate, move up one step on the next test. Jumping straight to a high setting “to be safe” produces bruising and callus thickening, which then forces even higher settings — a self-reinforcing loop. Once a patient finds their working setting, it belongs in their monitoring record so substitutes do not guess.
Alternate sites: ask before using them
Palm and forearm sampling works on some meters, but readings from alternate sites lag fingertip values during falling glucose — the exact situation where a lagged reading is most dangerous. Manufacturer approval and a healthcare provider’s say-so come before alternate-site testing, not after.
Steps 4 and 5: firing, and what to do with the drop
Press the capped device flat against the side of the fingertip — the palmar surface, slightly off centre, on the middle or ring finger. Firm contact matters: the depth dial assumes the cap is resting on skin, so a device held in mid-air fires shallower and a device pressed hard fires deeper. Press the release button.
After the puncture, let the drop form. Gentle proximal massage is acceptable; strong milking is not, because it pushes interstitial fluid into the sample and dilutes the reading. What to do with the first drop varies by protocol — North American practice generally wipes it, some testing systems require using it, and the safe instruction is to follow the protocol your test manufacturer specifies rather than habit. That conflict, and why both exist, is explained in our heel incision device comparison in the neonatal context.
Step 6: ejecting the lancet safely
Remove the cap. Most pens have an ejector — a slide or plunger that pushes the used lancet out over a sharps container. On devices without one, re-cap the used lancet with the protective cap you set aside in step 1, then pull it free and drop the whole thing into the container.
Do not leave the lancet in the pen between tests. Three reasons, all practical: a stored needle is an exposure hazard for whoever picks up the pen next; a used needle left seated contaminates the holder’s nose, which no weekly cleaning schedule can undo; and forgetting it is the direct route to firing a dull, contaminated needle the next morning. Eject every time, immediately.
Household sharps disposal follows local rules — some regions accept capped lancets in household waste through an approved container, others provide collection. The container rule is constant: rigid, puncture-resistant, closed before it is three-quarters full.
Cleaning and disinfecting: the schedule and the method
Regulators expect a reusable lancing device’s instructions for use to include a cleaning and disinfection procedure validated by the manufacturer — and they expect the user to follow it. The schedule below reflects published manufacturer guidance and is a reasonable default when a facility writes its own procedure.
| When | What | How |
|---|---|---|
| Weekly minimum, and whenever visible soil appears | Entire device surface and inside of the cap | Remove the lancet first. Wipe with a cloth moistened with 70% isopropanol or 70% ethanol. Repeat with a fresh cloth for disinfection. Air dry. |
| Before anyone else handles the device | As above | Same method. Then remind them the pen is not theirs to use. |
| Never | Openings, plunger slot, battery compartment | No liquid enters openings — moisture kills the mechanism and voids the design assumption behind the validated procedure. |
| Never | Soaking, autoclaving, ultrasonic baths | Not validated for these devices. They deform plastic and drive fluid inside. |
Watch for damage from repeated disinfection cycles: residue around buttons, stiffness when priming, difficulty inserting the lancet. Any of those means the pen retires. Slight discolouration after many cycles is cosmetic and expected.
The deeper point for buyers: a supplier who cannot produce the validation behind a cleaning procedure is selling a pen whose instructions were written by the marketing department. This is one of the checks in our manufacturer checklist that costs nothing to ask and filters hard.
How to use a lancing device safely: five recurring mistakes
These are the errors that appear repeatedly in incident reports and outbreak investigations, in roughly descending order of harm.
- Sharing the pen. Family members sharing one pen is the commonest form. The lancets are fresh; the holder is not. Hepatitis B survives dried blood on surfaces for about a week, which is far longer than a household cleaning interval.
- Assisted monitoring with a reusable pen. A carer firing the same pen for several patients transmits whatever is on the base. The correct tool is a single-use retractable safety lancet — one per patient, discarded intact.
- Leaving the lancet in the pen. Covered above. It contaminates the holder and hands the next user a spent needle.
- Cleaning by soaking. Submerging the device or running it through an ultrasonic bath drives fluid into the mechanism and is not covered by any validated procedure.
- Cranking the depth dial up permanently. High settings on thin skin produce bruising and scar tissue, which then need higher settings. Adjust to the patient, not to habit.
Buying note: what to check before a facility adopts a pen
For distributors and clinics, four lines in the instructions for use decide whether a lancing device can be adopted at all:
- “Single patient use only” marked on the device and in the IFU. FDA guidance expects it explicitly.
- A validated cleaning and disinfection procedure, named products and contact details included.
- An explicit prohibition on use for assisted blood draws — the IFU must say the pen is not to be used by a carer on multiple patients, with the single-use retractable alternative named.
- Named compatible lancet formats with gauge and depth stated together, so reorders do not substitute silently. Depth and gauge are independent variables; a “28G lancet” alone does not specify a product.
Devices we supply feed this category as well as the single-use one: the twist-top lancets that most pens take, in 14G to 33G with gamma sterilisation and lot-level traceability. Where the plan is to move a facility from reusable pens to single-use retractable devices — the direction US guidance and outbreak history both point — our safety lancet versus lancing device comparison sets out the trade-offs honestly, including what the pen still does better.
Frequently asked questions
How often should you change the lancet in a lancing device?
Every test, without exception. The needle is sterile once. After one puncture it is contaminated and blunted, and reusing it hurts more and carries infection risk. Anyone who finds themselves “just one more use” away from convenience is describing a habit, not a protocol.
Can family members share a lancing device if everyone uses their own lancets?
No. The reusable holder contacts blood and is not cleanable to a shared standard. Regulators classify it as a single-patient device, and outbreak investigations back the position with transmission data.
How do you set the depth on an adjustable lancing device?
Turn the dial to the lowest setting, fire, and move up one step per test until the drop is adequate. Skin thickness and the blood volume the test needs decide the setting — not how deep you press, which only adds to the depth and bruises the site.
Why did my lancing device stop working after cleaning?
Likely liquid reached an opening. These devices are wiped, never soaked; moisture in the plunger slot corrodes or gums the spring. If priming feels stiff or the release sticks, retire the device — it is not repairable at point of care.
Is a lancing device better than a single-use safety lancet?
For one person testing several times a day at home, yes: cheaper per test, adjustable, less waste. For anything involving more than one patient — wards, screening, care homes — the single-use retractable lancet is the only defensible choice, and in the US the multi-patient pen has no lawful route. Our comparison article works through both cases.
What gauge lancet should a lancing device use?
30G at 1.5 mm depth covers routine glucose monitoring with the least trauma. Move to 28G or 26G for thicker skin, and to 21G–23G only for tests needing larger volumes. The gauge and depth chart gives the full matrix with the reasoning.
Need lancets that feed your devices — or replace them?
We supply both sides of this decision. Free samples are available: twist-top lancets in 14G–33G for pen-based programmes, or single-use retractable safety lancets from 30G/1.5 mm to 17G blade for facilities moving to single-patient sampling. Tell us the device model in service, the gauge and depth in use, the destination market and annual volume, and we will send samples with the specification annex, lot-level sterilisation certificate, and a quotation stating volume tier, incoterm, validity and lead time together.
Shandong Lianfa has manufactured capillary blood collection devices since 1992, on a 20,000 m² campus with 10,000 m² of ISO Class 8 and 2,000 m² of ISO Class 7 cleanrooms and in-house biological, chemical and physical laboratories. We hold ISO 9001, ISO 13485:2016, CE (MDR), UKCA, FDA 510(k), NMPA GMP and KFDA; about 80% of output is exported to the United States, Brazil, Canada and Europe, and private-label programmes run from our own lines.