Pressure-Activated Safety Lancet Review: Shandong Lianfa PA Series
This pressure activated safety lancet review covers the Shandong Lianfa PA series, eight stock references running from a 30G needle at 1.5 mm to a 17G blade at 2.0 mm. We make the device, so treat the enthusiasm with suspicion and test the claims instead. What follows is a teardown of the mechanism, three bench tests you can repeat in your own warehouse with a force gauge and a ruler, the full reference table, a comparison against BD Microtainer, Unistik 3 and Accu-Chek Safe-T-Pro on published specifications, and four situations where the PA is the wrong device to buy. That last section is the one procurement teams should read first.
Short answer: the PA is a pressure-activated, single-use safety lancet with a tri-bevel needle, dual-spring firing, self-destructing retraction and eight SKUs covering 1.5–2.2 mm. It suits adult fingerstick at moderate to high volume, distributor and OEM supply. It is unsuitable for neonatal heel sampling, for anyone needing depths below 1.5 mm, and for purchasers who want a continuous depth dial.
Contents
- What this pressure activated safety lancet review covers
- Design teardown: what is inside a PA lancet
- Activation: what pressure-activated means, and how to measure it
- Depth consistency: the number on the box versus the hole in the finger
- The full range: eight references from 30G to a 17G blade
- Sterilisation, shelf life and traceability
- How the PA compares with three devices you already stock
- Where the PA fits, and four places it does not
- Who should buy it, and five questions to ask first
- Frequently asked questions
What this pressure activated safety lancet review covers
A review written by the manufacturer has one structural problem: the author benefits from a favourable conclusion. The only honest way around it is to make the review falsifiable. Every claim below either comes from a published specification you can check on someone else’s website, or from a test you can run yourself in under an hour.
So the scope is narrow and mechanical. We look at what the device is made of, how it fires, how deep it actually goes, what the range covers, and how it sits against three competitor devices on the specifications their makers publish. We do not score it out of ten. We do not name it the best. If you want a ranking, this is not that article.
How the device was assessed
The assessment used four inputs. First, dimensional checks on stock samples: needle gauge against ISO 9626 nominal outside diameter, and penetration depth against the figure printed on the carton. Second, a firing test on twenty units from one lot, measuring peak activation force and retraction completeness. Third, a review of the technical file: needle material, spring specification, latch geometry, sterilisation validation and lot coding. Fourth, a side-by-side comparison with published specifications for BD Microtainer Contact-Activated, Owen Mumford Unistik 3 and Accu-Chek Safe-T-Pro.
Twenty units is not a clinical trial. It is enough to catch a lot that is drifting, which is the failure mode that actually reaches customers. If a supplier cannot survive that test, the conversation ends there regardless of price.
What this review does not claim
No clinical data is presented here. We have not run a pain study, and we would be sceptical of any lancet maker who quotes one without naming the ethics approval and the sample size. Pain in fingerstick is dominated by depth, site selection and operator technique, in roughly that order. A needle geometry can shave a little off the top. It cannot rescue a 2.2 mm puncture on a thin finger.
We also do not claim the PA is better than the devices in the comparison table. They are different products aimed at different buyers, and in two respects covered later, they are plainly better than ours.
Design teardown: what is inside a PA lancet
Strip the housing off and a safety lancet is four parts: a needle, a spring or two, a latch, and a body that has to do three unrelated jobs at once. Hold the needle, shield it before and after use, and give the operator something to grip that does not rotate under pressure.
Most of the cost difference between a cheap lancet and an expensive one lives in the latch and the springs, not the needle. The needle is drawn wire. The latch is where a device either self-destructs reliably or quietly becomes reusable.
The needle: tri-bevel geometry and the third facet
A standard hypodermic needle has a single bevel: one angled plane ground across the tip. A tri-bevel needle has three, cut at different angles, meeting at a point that is closer to a lancet blade than to a syringe.
The practical effect is entry resistance. A single-bevel tip presents a broad cutting edge perpendicular to the tissue and pushes a wedge of skin ahead of it. Three facets distribute the cut so the tip parts tissue instead of compressing it. On a fingertip, where the epidermis is thick and packed with nerve endings, that difference is felt.
What the geometry cannot do is compensate for excessive depth. The papillary dermis, where the capillary loops that actually bleed sit, lies roughly 1.0 to 1.5 mm below the surface of the fingertip. Total epidermis plus dermis is about 1.5 to 2.0 mm. Past 1.5 mm you are cutting into tissue that adds pain and bruising but very little volume. This is why the 30G at 1.5 mm is the default for routine glucose monitoring, and why the deep references exist for a narrower set of indications.
The dual-spring mechanism
Cheap safety lancets use one spring to drive the needle forward, and rely on the same spring’s rebound or a separate light spring to pull it back. The failure is well known: if forward drive and retraction share an energy budget, a marginal lot retracts slowly or not completely, and a partially exposed tip is exactly the hazard the device was bought to eliminate.
The PA separates the two. One spring drives, a second retracts. That makes the retraction stroke independent of how much energy the forward stroke consumed, which matters when the device is fired against a calloused fingertip rather than a test rig.
Two springs cost more than one. They also add an assembly step that has to be controlled, because a mis-seated retraction spring produces a device that fires correctly and then fails to withdraw. When you audit a supplier, this is one of the few internal details you can probe without opening the device: ask for the retraction force specification and the acceptance range at end-of-line. A supplier with a number will answer in a sentence. A supplier without one will describe the spring as “high quality”.
The self-destruct latch
US special controls for this device type require a mechanism that makes reuse impossible, and require mechanical testing to demonstrate it. The PA does this with a latch that releases once. After firing, the needle retracts and the latch cannot re-engage, so a second push does nothing.
Test it yourself, and test it hard. Fire a unit, then push the activation face against a firm surface with roughly twice the force it took to fire the first time, and watch whether anything moves. Repeat on five units. A latch that survives the first fire but yields on a hard second push is a design defect that will show up in a ward where someone is in a hurry.
The reversible cap, and what it does not do
This is the feature most often misread, including by people selling it. The PA has a patented cap that can be flipped back over the tip after use, so a used device can be carried to a sharps container without an exposed needle.
That is a handling improvement, not the primary safety mechanism. The primary mechanism is retraction. Once the device has fired and retracted, the tip is already inside the housing and the cap is a second layer. What the cap genuinely solves is the awkward moment between the puncture and the sharps bin, when a device is lying on a tray with a contaminated tip and someone has to pick it up.
It does not protect the operator during the puncture itself. Nothing on a lancet can, because the finger is being cut by definition. Claims that a recapping feature prevents needlestick injuries overstate what a cap can do; the engineering control is the retraction, and US OSHA’s bloodborne pathogens standard looks for engineering controls, not careful habits. We would rather a buyer understood the distinction than repeated our marketing line back to an auditor.
Activation: what pressure-activated means, and how to measure it
Pressure activation means the device fires when its front face is pressed against the skin hard enough to release the latch. There is no button and no separate arming step. The force comes from the operator’s hand, transmitted through the device body.
This design has one clear advantage and one clear cost. The advantage is that the device cannot fire in mid-air. It has to be seated on a surface, which removes the commonest cause of accidental discharge in a busy clinic: a device triggered while being unpacked or passed between hands. The cost is that depth and force are coupled. Press harder, and you compress the tissue further before the latch releases, so the effective penetration is deeper than the nominal figure.
That coupling is the single most important thing to understand about any pressure-activated lancet, ours included. It is also why the nominal depth on the carton is a design value measured under a defined test condition, not a promise about what happens on a specific finger.
A bench test any buyer can run
You do not need a laboratory. A push-pull force gauge with a flat probe, a rigid stand, and twenty units from one lot will do.
- Clamp the device vertically with the activation face down.
- Lower the probe at a steady rate until the device fires. Record the peak force.
- Repeat on all twenty. Note the mean, the minimum and the maximum.
- Fire each unit once more against a rigid block and confirm no needle travels.
- Check retraction by looking into the aperture under a bright light, or by pressing the tip against a sheet of paper and confirming no perforation.
What you are looking for is not a particular number. It is spread. A lot with a tight distribution around a moderate force will feel the same in every operator’s hand. A lot with a wide spread produces the two complaints that fill complaint databases: “it would not fire” from a nurse with a light touch and “it fired too deep” from one who presses hard.
Ask the supplier for their specification before you test. If they quote a force range, you can check the lot against it. If they cannot, the test still tells you whether the lot is internally consistent, which is what matters for a repeat order.
Pressure-activated or push-button: which to pick
Push-button devices separate activation from seating. The operator places the device, then presses a button, and the force of the button press does not add to penetration depth. That gives a more repeatable depth and more control over timing.
The trade is a second action and a second failure point. Buttons get pressed early. On a struggling patient, a device that fires the instant it touches skin is often safer than one that requires a deliberate press while the hand is moving.
Our rule of thumb, and it is only a rule of thumb: pressure activation for high-volume adult screening where speed matters and the operator is trained; push-button where the patient is likely to move, or where depth repeatability across operators is the audit criterion. If you are choosing between the two architectures rather than between brands, the comparison is set out in our pressure-activated versus push-button guide.
Depth consistency: the number on the box versus the hole in the finger
Every lancet carton carries a depth. Treat that number as a design target with a tolerance, not a measurement of your patient.
The PA’s depths run 1.5, 1.8, 2.0 and 2.2 mm across the range. Each is set by the distance the needle tip protrudes past the front face at full extension, limited by an internal stop. Two things then move the real number: how far the front face sinks into the skin before the latch releases, and how thick the skin at the chosen site happens to be.
Why 1.8 mm on a calloused finger is not 1.8 mm
Fingertip skin is not uniform. The palmar surface of a distal phalanx carries a thick stratum corneum, and in people who work with their hands, that layer can be well over a millimetre. A device rated at 1.8 mm may only reach the deeper epidermis on such a finger, produce a poor drop, and get blamed for a bad lancet.
The reverse also happens. On a thin-skinned older patient, the same 1.8 mm reaches the full papillary dermis easily, and if the operator presses while firing, it goes further than intended. Bruising follows.
This is not a defect in the device. It is the reason a range exists and the reason site selection is taught. The fifth finger should not be used at all: the distance from skin to bone is too small, and the bone is close enough that depth limits stop protecting you. The middle and ring fingers, on the palmar surface slightly off centre, are the standard for a reason.
The depth that actually produces a drop
Here is the part that changes how you read a specification sheet. The capillary loops that supply a fingertip sit in the papillary dermis, roughly 1.0 to 1.5 mm below the surface. Reaching them produces a free-flowing drop. Going deeper does not produce proportionally more blood; it cuts more tissue, which means more pain, more bruising and a higher chance of squeezing the site, which dilutes the sample with interstitial fluid.
Put that beside the blood volumes that modern tests actually need. A glucose meter takes 0.5 to 1 microlitre. An HbA1c cartridge takes around 2 microlitres. A CRP cartridge takes about 12, and a lipid panel about 19. One drop from a fingertip is commonly estimated at 30 to 50 microlitres. For most point-of-care tests, a single drop is not a constraint; it is a surplus.
Which means the deep end of the PA range is not the “better” end. The 30G at 1.5 mm is the right default for routine capillary glucose. The 21G at 2.2 mm is for thick, calloused skin, cold patients with poor perfusion, or tests demanding a larger volume in one go. Buying the deepest reference because it sounds more capable is a common procurement error, and it shows up later as bruising complaints and refused repeat tests.
Depth limits by patient, not by preference. Adults: no more than 2.4 mm at the fingertip. Infants under one year: heel only, no more than 2.0 mm. Preterm and low-birth-weight infants: 0.65 to 0.85 mm. These limits come from the distance between skin and bone, and exceeding them risks bone or periosteum injury. A 2.2 mm adult lancet is never a neonatal device, whatever the sales sheet implies.
The full range: eight references from 30G to a 17G blade
The PA is not one product. It is eight stock references, and the two dimensions that define them — needle gauge and penetration depth — vary independently. That independence is the most useful thing in this table.
| Ref | Tip | Nominal OD (ISO 9626) | Depth | Typical use |
|---|---|---|---|---|
| 04-3015 | 30G needle | 0.30 mm | 1.5 mm | Routine capillary glucose, frequent testing, thin skin |
| 04-2818 | 28G needle | 0.36 mm | 1.8 mm | General adult fingerstick, HbA1c |
| 04-2618 | 26G needle | 0.45 mm | 1.8 mm | General adult fingerstick where flow matters more than comfort |
| 04-2118 | 21G needle | 0.80 mm | 1.8 mm | Poor perfusion, thicker skin, same depth as 26G with more flow |
| 04-2322 | 23G needle | 0.60 mm | 2.2 mm | Calloused fingers, larger volume, cholesterol and chemistry panels |
| 04-2122 | 21G needle | 0.80 mm | 2.2 mm | Maximum flow at maximum safe adult depth |
| 04-1720 | 17G blade | n/a | 2.0 mm | Very high flow, dense callus, thick finger pads |
| 04-1822 | 18G blade | n/a | 2.2 mm | Highest flow in the range, tough skin, large-volume draws |
Packaging is 50 pieces per box and 2,000 per case, which is forty boxes to a case. That ratio is worth keeping in your head, because small-clinic buyers routinely order in boxes and distributors order in cases, and the per-unit price gap between the two is a packaging and handling cost, not a manufacturing one.
This is also where a pressure activated safety lancet review earns its keep for procurement. The reference matrix, not the mechanism, decides whether one supplier can cover a whole protocol with a single part number family and a single set of documents.
Why the same 21G appears twice
Look at 04-2118 and 04-2122. Both are 21G. One is 1.8 mm, the other 2.2 mm. This single pair dismantles the most common simplification in lancet procurement: the belief that gauge determines depth.
It does not. Gauge determines the diameter of the hole. Depth determines how far the hole goes. A 21G at 1.8 mm and a 26G at 1.8 mm cut to the same depth; the 21G produces a wider channel and more flow. A 21G at 2.2 mm produces both more flow and more tissue damage than the same needle at 1.8 mm.
Practically, this means a specification of “21G lancet” is incomplete, and a quotation built on it is not comparable to another quotation built on the same phrase. Ask for gauge and depth, in millimetres, together. We have seen tender documents specify a gauge and leave depth to the bidder, which guarantees the bids describe different products.
The blade versions, and the neonatal trap
The two blade references exist for a specific problem: a needle can be too fine to produce adequate flow from heavily calloused skin, no matter how deep it goes. A blade cuts a short line rather than piercing a round hole, exposing more capillary length per unit of penetration. At 17G and 18G, these are the highest-flow devices in the range.
They are also the ones most likely to be mis-sold. A blade at 2.0 or 2.2 mm looks, on a spec sheet, like a neonatal heel device. It is not. Neonatal heel sampling is limited to 2.0 mm for term infants and 0.65 to 0.85 mm for preterm infants, and the geometry that matters there is a shallow, wide incision made with a device designed to sit against a heel. A 2.2 mm adult blade applied to a newborn heel can reach the calcaneus.
We make a separate heel blade range for this reason, at 0.65 × 1.40, 0.85 × 1.75, 1.00 × 2.50 and 1.40 × 2.75 mm. If a supplier offers you one lancet for both adult fingers and newborn heels, that is not versatility. It is a product line that has not been thought through, and it is a question worth asking on the first call. Our neonatal heel incision device guide covers the depth logic in full.
Sterilisation, shelf life and traceability
Three things decide whether a sterile lancet is sterile at the moment of use: the sterilisation process, the barrier that keeps it sterile, and whether you can trace the unit in your hand back to a validated batch.
Ethylene oxide or gamma
Both are used for this product type, and both are acceptable when validated. The differences matter commercially more than clinically. Gamma is a terminal process done at a contract irradiation facility, priced by volume and dose, and it penetrates sealed packaging without the need for a breathable sterile barrier. Ethylene oxide needs a permeable barrier, a controlled cycle, and aeration time to drive off residue, which makes it slower and adds a release step.
Whichever route a supplier uses, the evidence you want is the same: a validated dose, a dose audit programme on a defined frequency, and a certificate per batch stating the actual dose range achieved. A line on a quotation that says “gamma sterilised” with no dose is a claim about a process, not evidence that the process happened to your goods.
Lot coding and what to demand
The PA carries a lot code that ties the carton to a manufacturing and sterilisation batch. That code is the entry point for everything an auditor or a complaint investigation will need: which needle lot, which assembly run, which sterilisation cycle, which release test results.
When a complaint arrives — and in this category complaints are usually “did not retract” or “fired too early” — the lot code decides whether the answer is a single unit, a batch, or a design change. In FDA’s public device complaint database, a recurring pattern for this product code is a report of a safety mechanism failing where the physical device was never returned, so the investigation closes as unconfirmed. Traceability is what stops a single unverified report from becoming an unanswerable one.
Ask for three documents with the first shipment: the lot-level certificate of analysis, the sterilisation certificate for that lot, and the shelf-life validation summary that supports the expiry printed on the box. A supplier who has these will send them without negotiation. A supplier who has to “check with the factory” for two weeks is telling you something about the file.
How the PA compares with three devices you already stock
A pressure activated safety lancet review that compares against unnamed competitors is a brochure with a table in it. These three are the devices most likely to be on the shelf next to ours in a European or North American clinic, and all three publish enough specification to compare honestly.
| Shandong Lianfa PA | BD Microtainer Contact-Activated | Unistik 3 (Owen Mumford) | Accu-Chek Safe-T-Pro | |
|---|---|---|---|---|
| Activation | Pressure (front face) | Contact pressure | Side button with alignment guide | Pressure |
| Depth options | 1.5 / 1.8 / 2.0 / 2.2 mm | 1.5 / 1.8 mm needles, 2.0 mm blade | 1.5 / 1.8 / 2.0 mm across five variants | 1.5 mm fixed (Uno); multi-depth (Plus) |
| Gauge coverage | 30G to 21G, plus 17G and 18G blades | 30G and 21G, plus 1.5 × 2.0 mm blade | 30G to 18G | Not fully published |
| Retraction | Self-destructing, dual spring | Automatic after puncture | Retracts after use | Needle fully consumed after use |
| Traceability | Lot code on carton | Lot number laser-etched on each lancet | Lot number label | Batch on carton |
| Pack format | 50 per box, 2,000 per case | 200 per box | 100 or 200 per box | 200 per box |
| Supply route | Factory direct, OEM and private label | Authorised distributors | Authorised distributors | Authorised distributors |
Two rows deserve comment, because they are where we lose.
Where the brands are genuinely better
The first is per-unit traceability. BD laser-etches the lot number onto each individual lancet. The PA carries its lot code on the carton. If a single device arrives at an investigation separated from its packaging, BD’s unit still identifies itself and ours does not. For a hospital with a rigorous incident-reporting process, that is a real advantage and we do not pretend otherwise. If unit-level traceability is in your specification, ask us for it explicitly and expect a different price and a longer lead time — but do not assume it is there by default.
The second is recognition. A ward nurse who has used BD Microtainer for fifteen years will not read a datasheet. Brand familiarity has genuine safety value in an environment with high staff turnover, and no specification table captures it. Distributors who switch a hospital to a factory-direct lancet without a training session are trading a small unit-cost saving for a real training cost, and that trade is often a bad one.
Where the factory-direct model wins is different: breadth of SKUs (eight references, including two blade sizes), private-label packaging from modest volumes, and the ability to hold a specification stable across repeat orders rather than substitute an equivalent when a line changes. If your requirement is “the same device, in our packaging, for three years”, that is our territory. If your requirement is “the brand our nurses already know”, it is not, and we will say so.
Where the PA fits, and four places it does not
A safety lancet is a narrow instrument. It makes one kind of hole in one kind of tissue. Most bad purchasing decisions in this category come from stretching it.
Four cases where the PA is the wrong choice
- Neonatal or infant heel sampling. The shallowest PA reference is 1.5 mm, and every heel protocol for infants caps at 2.0 mm with preterm limits far below that. Heel sampling needs a heel device with an incision geometry designed for a heel. Using a 30G adult lancet on a newborn is a protocol violation waiting to be written up. See our heel incision device comparison.
- Depths below 1.5 mm. Paediatric fingertips, thin skin, or any protocol calling for 0.8 to 1.2 mm. The PA range does not go there. The adjustable APA starts at 1.0 mm and is the right family for those cases.
- Anyone who needs a dial. If the requirement is one device covering several depths, a fixed-depth lancet is the wrong architecture no matter how many SKUs exist. That is what the adjustable safety lancet range is for, and the honest limitation there is that its positions are discrete steps, not continuous adjustment.
- Reusable-device workflows. A clinic that wants one handle and a box of disposable tips is buying a lancing device, not a safety lancet. Those are different products with different regulatory positions, and a multi-patient reusable lancing device has no lawful US route at all. Our safety lancet versus lancing device article explains why.
Where it does fit
Adult fingerstick at volume, where the operator is trained and the patient is an adult: that is the core case. Pharmacies running point-of-care panels, occupational health screening, blood donation pre-screening, diabetes clinics, and distributors supplying those settings.
It also fits OEM and private-label programmes, which is a different kind of fit. A distributor with an established brand and a stable annual volume can specify the reference mix, the box artwork and the languages on the label, and hold that specification across reorders. That is not a product advantage; it is a supply-chain arrangement, and it is the reason a factory-direct conversation is worth having even when the device itself is unremarkable.
Who should buy it, and five questions to ask first
Three buyer profiles get value from this device, and one does not.
The four exclusions above are the working output of this pressure activated safety lancet review. They decide more deals than any row in a specification table, because a device that is wrong for the protocol will fail in use regardless of how well it is made.
Distributors supplying clinics and pharmacies, who need a broad gauge and depth matrix from one supplier and want private-label packaging. The eight-reference range is the argument here, not any single SKU.
Point-of-care operators running screening programmes, where cost per test matters and the device is one line in a protocol that is already written. The 30G at 1.5 mm will cover most of that volume.
OEM buyers building a own-brand range, who need a manufacturer able to hold a specification and produce documentation per batch. The relevant question is not “what does your lancet do” but “what happens when I reorder in eighteen months”.
The buyer who should walk away is anyone whose requirement is brand recognition in a specific hospital account, or unit-level traceability, or a depth below 1.5 mm. In each of those cases a competitor or a different product family is the better answer, and a supplier who tells you otherwise is optimizing for the order rather than the relationship.
Five questions that separate a factory from a trader
- What is the retraction force specification, and what is the end-of-line acceptance range?
- What dose range was validated for sterilisation, and can you send the certificate for the lot you are quoting?
- Which references share a needle gauge but differ in depth, and why does your catalogue carry both?
- What is the shelf-life validation basis for the expiry you print?
- If I reorder the same reference in eighteen months, what is your change-control notification commitment?
None of these can be answered with a brochure sentence. That is the point. Question three is the one that most reliably separates a supplier who understands the product from one who is reselling a price list, because the answer requires knowing that gauge and depth are independent variables.
For the wider supplier assessment, including the documents to request and the audit points that matter, see our 12-point safety lancet manufacturer checklist and the safety lancet FAQ.
Frequently asked questions
Is a pressure activated safety lancet review enough to choose a supplier?
No. A review tells you what a device is and how it behaves. Choosing a supplier also requires the quality system, the registration status for your market, the change-control history and a sample test on your own bench. Use a review to build a shortlist, then test.
What does pressure activated mean on a safety lancet?
The device fires when its front face is pressed against the skin with enough force to release an internal latch. There is no button. It cannot fire in mid-air, which reduces accidental discharge, but it also couples operator force to penetration depth.
Can the PA be used on newborns?
No. The shallowest reference is 1.5 mm and the range goes to 2.2 mm, while infant heel protocols cap at 2.0 mm for term infants and 0.65 to 0.85 mm for preterm infants. Neonatal sampling requires a dedicated heel incision device.
Why does Shandong Lianfa make both 21G at 1.8 mm and 21G at 2.2 mm?
Because gauge and depth answer different questions. Gauge sets the channel width and therefore the flow; depth sets how far into the tissue the channel goes. A 21G at 1.8 mm suits a patient who needs more flow than a 26G but not a deeper puncture. A 21G at 2.2 mm is for thick skin or large-volume draws.
Does the recapping cap prevent needlestick injuries?
It reduces exposure during handling and disposal, which is useful. It does not protect the operator during the puncture, and it is not the engineering control. The retraction mechanism is. Regulators look for an integrated safety feature, not a cap.
Can I get the PA with private-label packaging?
Yes. Box artwork, label languages, reference mix and carton configuration can be specified, subject to minimum quantities that are higher than for standard stock. Ask for the sample and the artwork proof before the purchase order, and confirm in writing which regulatory registrations the private-label version will carry.
Test the PA on your own bench
We send free samples — twenty pieces of the reference you are considering — so your force test, fit check and depth measurement start with the real device rather than a datasheet. Tell us the gauge, depth, market and annual volume, and we will send the samples with the specification annex, the sterilisation certificate for the lot, and a quotation that states volume tier, incoterm, validity and lead time together.
Shandong Lianfa has manufactured capillary blood collection devices since 1992, on a 20,000 m² campus with 10,000 m² of ISO Class 8 and 2,000 m² of ISO Class 7 cleanrooms, plus in-house biological, chemical and physical laboratories. We hold ISO 9001, ISO 13485:2016, CE (MDR), UKCA, FDA 510(k), NMPA GMP and KFDA, and roughly 80% of output is exported to the United States, Brazil, Canada and Europe. Private-label and OEM programmes run from our own lines.
Learn more about Lianfa pressure-activated safety lancets.